Covid-19 screening: Please may I take your temperature?

anton-SZ1DDwCPqkE-unsplash-1

Question:  

Does temperature screening detect Covid-19 infected patients?

Bottom line:

Even though handheld infrared thermometers are convenient to use to check if a patient has an elevated body temperature, they aren’t sufficiently accurate for screening purposes. Using recent Office of National Statistics data on the prevalence of Covid-19 in the population the false positive rate is too high (>95%). The major confounders regarding accuracy are environmental temperature, humidity, gender, exercise, and age.

Background

In the recent Cochrane ‘Recommendations for the re-opening of dental services: a rapid review of international sources’ (Cochrane, 2020) some of the guidelines recommended temperature screening of the patients at reception for elevated body temperature. The rational being that if a patient is infected with Covid -19 the body’s response to the virus often results in core body temperature increase. Traditionally body temperature was taken with a glass/mercury or electronic thermometer that required intimate contact with the patient but now there are handheld infrared thermometers (HIRT) that are quick and require only skin contact via the ear canal, or contactless by measuring the forehead skin temperature. In this opinion paper we aim to find out how effective temperature screening is in detecting Covid infected patients.

Methods

To reduced unnecessary searching through the literature to answer this question the diagnostic accuracy data was extracted from two rapid reviews, the most recent from the Emergency Care Research Institute (ECRI, 2020), and the second from the Canadian Agency for Drugs and Technologies in Health (CADTH, 2014). It was possible to extract the sensitivity and specificity data for 10 studies measuring the effectiveness of HIRT for forehead temperature (FT), and 7 studies measuring ear (tympanic) temperature (TT). The data was extracted and back transformed into a classic 2×2 table giving us the true positive (TP), false negative (FN), false positive (FP) and true negative (TN) data and meta-analysis was carried out using the ‘mada’ package in R. The summary estimate for sensitivity and specificity for TT and FT are tabulated below (See Table 1).

Table 1. Summary estimates for IR thermometer

Measurement location Sensitivity (95%CI) Specificity (95%CI)
Tympanic temperature 78.7 (69.4 to 85.8) 91.8 (75.7 to 97.6)
Forehead temperature 51.1 (19.3 to 82.0) 97.1 (92.2 to 99.0)

The results for the TT and TF results were plotted together on to a Summary Receiver Operating Characteristic (sROC) curve for comparison. The y-axis represents the sensitivity (1.0 =100%), and the x-axis represents 1- specificity (0.1 = 10%), the solid triangle and circle are the summary estimates, and the ellipses are the 95% confidence areas. ( Figure 1.)

Figure 1. Comparison of diagnostic accuracy tests

ROCtemp

A perfect diagnostic test would be in the extreme top left corner representing 100% true positives and 0% false positives and from the chart we can see that point is outside the 95% confidence area meaning that both tests are poor for screening. To clarify this point I will transform the sensitivity/specificity results into a frequency tree (Figure 2) using a diagnostic test calculator.

Figure 2. Frequency trees for diagnostic tests for screening Covid-19 (Prevalence 1:400)

Frequency tree

If these thermometers are used for screening patient for Covid infection then out of every  838 patients who test positive with elevated TT only 20 will be infected which corresponds to a probability of 2.4% and for the FT that rises to 4.2%.

What happens if we use these thermometers to confirm a diagnosis of fever where we set a prevalence of 95% instead of 0.0025% (Figure 3.).

Figure 3. Frequency trees for diagnostic tests to confirm fever

Frequency tree_v1

When these devices are used to confirm whether the patient has a true fever the TT test is 78.9% accurate and the FT is only 50% accurate.

Conclusions

The authors of the  recent ECRI report concluded:-

 Temperature screening programs using IR alone or with a questionnaire for mass screening are ineffective for detecting infected persons, based on our review of evidence from 2 large systematic reviews (SRs), 3 simulation studies, and 6 diagnostic cohort studies (not included in the SRs). Under best-case scenarios, simulation studies suggest such screening will miss more than half of infected individuals. They are ineffective for mass screening because of the low number of infected individuals who have fever at the time of screening and inconsistent technique by operators.

Comments

Both pieces of diagnostic equipment produce highly variable results whether used for screening or confirming a diagnosis of febrile illness. The limitations are well described in the CADTH report (CADTH, 2014):

The retrieved studies have mentioned potential confounders for measure of temperature such as sweat, gender, age, the range of temperature, the rater, physical activity, the use of antipyretic drugs and emotional state. These factors are even more susceptible to vary in a real world conditions than in a clinical study setting. Moreover, the different brand/model/mode of devices used make it difficult to draw general conclusions on a class of thermometers. Also, a fair number of pediatric studies were included in the present review, limiting the extrapolation of their results to a general population.

In conclusion if a patients temperature needs to be taken then tympanic temperature is more reliable than forehead temperature, however its use for screening in the practice creates another layer of complexity in the cross infect/record keeping process with little diagnostic value.

 

Disclaimer:  The article has not been peer-reviewed; it should not replace individual clinical judgement, and the sources cited should be checked. The views expressed in this commentary represent the views of the author and not necessarily those of the host institution. The views are not a substitute for professional advice.

References

CADTH. 2014. Non-Contact Thermometers for Detecting Fever: A Review of Clinical Effectiveness [Online].  [Accessed 18th May 2020 ].

COCHRANE. 2020. Recommendations for the re-opening of dental services: a rapid review of international sources [Online].[Accessed 18th May 2020 ].

ECRI. 2020. Infrared Temperature Screening to Identify Potentially Infected Staff or Visitors Presenting to Healthcare Facilities during Infectious Disease Outbreaks [Online]. [Accessed 18th May 2020 ].

Photo Credits

Photo by Anton on Unsplash
Share on Facebook Tweet this on Twitter Share on LinkedIn Share on Google+
Mark as read
Create a personal elf note about this blog
Profile photo of Mark-Steven Howe

Mark-Steven Howe

Mark qualified from Birmingham University in 1988. After 4 years in general practice he joined the Royal Air Force Dental Branch on a Short Service Commission serving in the UK and overseas gaining his Diploma and Membership exams in general dental surgery. On leaving the RAF Mark completed his Fellowship to the Faculty of General Dental Practice (FFGDP) and has recently completed an MSc in Evidence-Based Healthcare at the University of Oxford.  Mark has been a full time general dental practitioner at Broadway Dental Care, Worcestershire since 1998. He is a Fellowship assessor for the Faculty of General Dental Practitioners UK, and examiner at the Royal College of Surgeons, Edinburgh.

More posts - Website

Follow me here –